We use cookies to improve your online experience. By continuing browsing this website, we assume you agree our use of cookies.
Categories

LCD Display Mobile Ventilator Machine YSAV90B

Model: YSAV90B
LCD display mobile ventilator machine YSAV90B
  • ModelYSAV90B
  • BrandYSENMED
  • PriceGet Price
  • MOQ1
  • Ship FromFrom Guangzhou,China
  • ShippingAny shipping terms convenient for customers
  • -
    +
    Categories:Respirators / Ventilators
    Share:
    Description Get Price Review


    SPECIFICATIONS
    Physical specifications
    Display 10.4'' LCD Display
    Mode: pneumatically powered and electrically controlled system
    Trolley: Fitted with 4 nos anti-static rubber castors; two of which are lockable for braking and easy maneuverability with foot operated brake provisions
    Technical Specifications:
    Gas requirement: Medical oxygen and nitrous oxide with a pressure ranging from 280~600kpa
    safety valve <12.5 kPa
    Respiratory Rate 1 ~ 99 bpm
    gas concentration output 21% ~ 100% adjustable
    Maximum Minute ventilation ≥18L / min
    modes of ventilation A/C, SIPPV,IPPV,SIMV,PSIMV,CPAP,PCV,PEEP,SIGN,SPONT, MANUAL
    Ventilator compliance ≤30 Ml/kPa
    PEEP: 0 ~ 1.0 kPa
    Ptr: -10 ~ 10 hPa
    IMV frequency: 1~20 bpm
    I/E Ratio: 4:1~1:4, Has the inverse ratio ventilation
    Tidal Volume Setting Range: 50~1500ML     Display Range: 0~2000ml ;   resolution ratio: 10ml
    Inspiratory Plateau: 0~1s
    Maximum Safety Pressure ≤ 6.0kPa
    Sigh: one deep breath per 80 controlled respirations, the inspiration time is 1.5 times of the setting point
    Pressure limit range: 1.0~ 6.0kPa
    Airway pressure alarm: Audible and visual and with yellow and red color indicating Lower: 0.1 kPa ~ 6.0 kPa;   Upper: 0 ~ 5.0 kPa
    ±0.2 kPa
    Ventilation Volume alarm: ≥ 18L/ MIN
    Power Supply Alarm Ac/dc power supply are after failing to send out alarm immediately
    Alarm Time: keep >120s
    The airway pressure continues to be higher than 15 hPa ±1 hPa for 15s±1s, then the machine will raise an audible alarm, the pressure will be displayed in red and the continuous high pressure red alarm wording is displayed on the screen of the anesthetic respirator.
    Operating conditions
    Ambient temperature: 10 ~ 40oC
    Relative humidity: no higher than 80 %
    Atmospheric pressure: 860 hPa ~ 1060 hPa
    Power requirement: 100-120 Vac, 50/60 Hz; 
    Attention: the AC power supply used for the anesthesia machine must be well     grounded.
    Attention: the anesthesia machine used must be equipped with a carbon dioxide monitor complying with ISO 9918:1993, an oxygen monitor complying with ISO 7767:1997 and an expiratory gas volume monitor complying with 51.101.4.2 of Medical Electrical Equipment Part II: Special Requirements for Safety and Basic Performance of Anesthesia System.
    Storage
    Ambient temperature: -15oC ~ +50oC
    Relative humidity: no higher than 95 %
    Atmospheric pressure: 86 kPa ~ 106 kPa.
      It should be stored in a room without corrosive gas and well ventilated
    Package
    packaging box comply with the requirement of GB/T 15464
      Between the packaging box and product, soft material with appropriate thickness provided to prevent loosening and mutual friction during transportation
      Moisture protection and rain protection to ensure that product is protected from natural damage.
    Safty & Alarm
    Gas source pressure Immediate alarm; “high level alarm” indicator flickers raising an audible alarm.
    Airway Pressure Alarm
    Write down your demand below and we will contact you within 24h!

    1/3
    X