  • ModelYSAV70A
  • BrandYSENMED
  • PriceGet Price
  • MOQ1
  • Ship FromFrom Guangzhou,China
  • ShippingAny shipping terms convenient for customers
  • -
    +
    Categories:Respirators / Ventilators
    Share:
    Description Get Price Review

    TECHNIQUES SPECIFICATIONS
    Physical specifications 
    Display  10.4'' LCD Display
    Mode: pneumatically powered and electrically controlled system
    Trolley:  Fitted with 4 nos anti-static rubber castors; two of which 
    are lockable for braking and easy maneuverability with 
    foot operated brake provisions
    Technical Specifications:
    Gas requirement:  Medical oxygen and nitrous oxide with a pressure ranging 
    from 280~600kpa
    safety valve  <12.5  kPa
    Respiratory Rate 1 ~ 99 bpm
    Oxygen concentration output 21% ~ 100%  adjustable
    Maximum Minute ventilation 18L / min
    modes of ventilation A/C, IPPV, SIPPV, IMV, SIMV, PEET. MANUAL, SIGH, 
    SPONT
    Ventilator compliance 30 Ml/kPa
    PEEP: 0 ~ 1.0 kPa
    Ptr:  -10 ~ 10 hPa
    IMV frequency: 1~20 bpm
    I/E Ratio: 4:1~1:4, Has the inverse ratio ventilation
    Tidal Volume Setting Range:  50~1500ML   Display Range: 0~2000ml ;  
    resolution ratio: 10ml
    Inspiratory Plateau: 0~1s
    Maximum Safety  Pressure ≤ 6.0kPa
    Sigh:  one deep breath per 80 controlled respirations, the inspiration
     time is 1.5 times of the setting point
    Pressure limit range: 1.0~ 6.0kPa
    Airway pressure alarm: Audible 
    and visual and with 
    yellow and red color indicating
    Lower: 0.1 kPa ~ 6.0 kPa;   Upper: 0 ~ 5.0  kPa
    ±0.2 kPa
    Ventilation Volume alarm:  ≥ 18L/ MIN 
    Power Supply Alarm Ac/dc power supply are after failing to send out alarm 
    immediately
    Alarm Time: keep >120s
    The airway pressure continues to be higher than 
    15 hPa ±1 hPa for 15s±1s, then the machine will raise an audible alarm, the pressure will 
    be displayed in red and the continuous high pressure red alarm wording is displayed 
    on the screen of the anesthetic respirator.
    Operating conditions
    Ambient temperature:  10 ~ 40oC
    Relative humidity: no higher than 80 %
    Atmospheric pressure: 860 hPa 1060 hPa
    Power requirement:  100-120 Vac, 50/60 Hz;  
    Attention: the AC power supply used for the anesthesia 
    machine must be well   grounded.
    Attention: the anesthesia machine used must be equipped 
    with a carbon dioxide monitor complying with ISO 9918:1993, 
    an oxygen monitor complying with ISO 7767
    :1997 and an 
    expiratory gas volume monitor complying with 51.101.4.2 of 
    Medical Electrical Equipment Part II: Special Requirements for 
    Safety and Basic Performance of Anesthesia System. 
    Storage
    Ambient temperature:  -15oC ~ +50oC
    Relative humidity:  no higher than 95 %
    Atmospheric pressure:  86 kPa ~ 106 kPa.
      It should be stored in a room without corrosive gas and well 
    ventilated
    Package
    packaging box comply with the requirement of GB/T 15464
      Between the packaging box and product, soft material with 
    appropriate thickness provided to prevent loosening and mutual 
    friction during transportation
      Moisture protection and rain protection to ensure that product is 
    protected from natural damage. 
    Safty & Alarm
    Gas source pressure Immediate alarm; “high level alarm” indicator flickers raising 
    an audible alarm.
    Airway Pressure Alarm The airway pressure exceeds 15hpa. The maximum delay is no 
    longer than 17s. The “high level alarm” indicator flickers raising 
    an audible alarm. 
    Tidal Volume Alarm Lower than 50ml, The alarm will be raised after one breathing 
    cycle. 
    Ventilation Volume Alarm Lower: 0~12L/Min;   Up: 18L/ Min
    Battery Voltage Alarm It alrms when the battery voltage is lower than 10.5V.
    Power On Self Test Self test before working, assure it's working normally. 
    STANDARD CONFIGURATIONS  
    QTY NAME
    1 set Main unit
    1 set Oxygen tube
    2 set Silicone Mask
    1 pc Power line
    4 pcs Threaded pipe
    2 pcs Standard Pipe
    1 set Oxygen Reducer
    1 set Humidifier
    1 pc Battery
    2 set flow sensor
    2 pcs A leather bag ( Blue)
    1 set Tools with the machine
    1 set User Manual (English Version)

     