SPECIFICATIONS | |
SPO2 performance | |
Range | SpO2: 0 to 100% PR: 25 to 250 bpm |
Resolution | SPO2: 1% PR: 1BPM |
Saturation Accuracy | Saturation: 70 to 100 Normal: 2% Motion or Low Perfusion: 3% Saturation: <70% ,undefined Normal: ±2bpm Motion or Low Perfusion: ±3bpm |
ETCO2 performance(For main stream ETCO2 sensor) | |
Range: | ETCO2: 0 to 152mmHg RR: 2 to 150 bpm |
Resolution | CO2: 0.1mmHg RR: 1BPM |
Accuracy | CO2: ±2mmHg(0-40mmHg), <10% of reading(41-152mmHg) RR: ±1BPM |
Electrical | |
AC Power Requirement | 100 to 240VAC 47 to 63Hz |
Power Comsumption | 20VA |
Batteries | |
Type | Alkaline Ni-Mh Batteries |
Operating Time | >20Hours(SO2 only) or >5Hours(ETCO2 and SpO2) |
Charge Time | <6hours<> |
Environment | |
Operating Temperature | 0 to 45 °C |
Storage Temperature | 20 to 60 °C |
Operating Humidity | 30 to 95%, (non-condensing) |
Operating Altitude | -500 to 5000 m |
Physical Characteristics | |
Dimensions | 337.5px*187.5px*70px |
Weight | 258g with batteries(not include SPO2 sensor and ETCO2 sensor) |
Alarms | Audible and visual alarms for high and low saturation 0 to 100% SPO2, pulse rate (25 to 250bpm), sensor off, 5 to 152 mmHg (ETCO2), 2 to 150BPM (RR), full storage and low battery Alarm Volume: 85dB |